34
8-6
Correcting The Internal Mass Value Of The GX-A series (Auto)
Calibrate referring to "8-3. Calibration Using An External weight".
This is method of correcting the internal mass weight value based on an external weight.
After calibration with the external mass, the balance automatically loads and unloads the
internal mass and corrects the internal mass value.
The available masses are as shown in the table below. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.
Setting procedure
1. The internal mass value cannot be corrected at factory settings.
Refer to "9. Function Switch and Initialization" and enable changing
of the function setting and correction the internal mass value.
2. In weighing mode, press and hold the SAMPLE key to
display
.
3. Press the SAMPLE key several times until appears.
4.
If does not display, perform Step 1.
5. Press the PRINT key to display
auto .
6. When preparation is completed, press the PRINT key.
7.
is displayed and the internal mass value is automatically corrected.
8. When adjustment of the internal mass value is completed, is
displayed and calibration is performed automatically with the adjusted internal
weight.
9. is displayed and you are returned to the weighing mode.
10. Place the external weight used for calibration on the balance check
whether the balance was corrected. If it is not corrected properly, return to
Step 2.
Model
Available mass
Factory
setting
Adjustable
range
GX-203A, GF-203A
50g, 100g, 200g
200g
-9.999g ~
+9.999g
GX-303A, GF-303A
50g, 100g ~ 300g (100g interval)
200g
GX-403A, GF-403A
50g, 100g ~ 400g (100g interval)
400g
GX-603A, GF-603A
50g, 100g ~ 600g (100g interval)
500g
GX-1003A, GF-1003A
50g, 100g ~ 1000g (100g interval)
1000g
GX-1603A, GF-1603A
50g, 100g ~ 1600g (100g interval)
1000g
GX-2002A, GF-2002A
500g, 1000g, 2000g
2000g
-99.99g ~
+99.99g
GX-3002A, GF-3002A
500g, 1000g ~ 3000g (1000g interval)
2000g
GX-4002A, GF-4002A
500g, 1000g ~ 4000g (1000g interval)
4000g
GX-6002A, GF-6002A
500g, 1000g ~ 6000g (1000g interval)
5000g
GX-10002A, GF-10002A 500g, 1000g ~ 10000g (1000g interval)
10000g
GX-6001A, GF-6001A
500g, 1000g ~ 6000g (1000g interval)
5000g
-999.9g ~
+999.9g
GX-10001A, GF-10001A 500g, 1000g ~ 10000g (1000g interval)
10000g
During
calibration
with
external
mass
Place
Press
several
times
Press
and
hold
x2