62
Transmitting all memory data at one time
Confirm that the "Serial interface (
5if
)" parameters are set properly.
Refer to "10. Function Table" and "Communication Manual " on the A&D website.
1. Press and hold the PRINT key until is displayed, then release the key.
2. Press the SAMPLE key to display
out
.
3. Press the PRINT key to display
out
with " " blinking.
4. Press the RE-ZERO key to display
out
with " " blinking.
5. Press the PRINT key to transmit all data using the RS-232C , USB.
6. The balance displays
when all data is transmitted.
Press the CAL key to return to the weighing mode.
Deleting all memory data at one time
1. Press and hold the PRINT key until
is displayed, then release the key.
2. Press the SAMPLE key several times to display
.
3. Press the PRINT key to display
with " " blinking.
4. Press the RE-ZERO key to display
with blinking.
5. Press the PRINT key to delete all data
6. The balance displays
end
and returns to the weighing mode.
12-2
Data Memory For Calibration And Calibration Test
Characteristic
□
Calibration data (when and how it is performed) and calibration test data can be
stored in memory.
□
All the data in memory is available to be output at one time to a printer or
personal computer.
Up to 50 data sets of the latest calibration or calibration test can be stored.
※
When the memory capacity has been reached to 50, "
fUl
" " "
illuminates in order in the upper left of the display as shown below.
Storing the calibration and calibration test data
1. Set the "Data memory (
data
)" parameter to "
data 2
". Refer to "10 Function Table".
2. With the settings above, each time calibration or calibration test is performed, the data is stored
automatically.
Upper left of the display