82
18.Programmable-Unit
This is a programmable unit conversion function. It multiplies the weighing data in grams by an arbitrary
coefficient set in the function table and displays the result.
The coefficient must be within the range between the minimum and maximum shown below. If the
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient setting
mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.
Model
Minimum
coefficient
Maximum coefficient
GX/GF-203A/303A/403A/603A/1003A/1603A
0.000001
1000
GX/GF-2002A/3002A/4002A/6002A/10002A
100
GX/GF-6001A/10001A
10
Operation
1 Press and hold the SAMPLE key until
ba5fnc
of the function table is displayed.
2 Press the SAMPLE key several times to display .
3 Press the PRINT key. The balance enters the mode to confirm or set the coefficient.
Confirming the coefficient
4 The current coefficient is displayed with the first digit blinking.
□
When it is not to be changed, press the CAL key and proceed to step 6.
□
When it is to be changed, press the RE-ZERO key and proceed
to step 5.
Setting the coefficient
5 Set the coefficient using the following keys.
SAMPLE key
To select a digit to change the value.
The selected digit blinks.
RE-ZERO key
To change the value.
MODE key
To change the decimal point position.
Each time the switch is pressed, the
decimal point position changes as
follows:
PRINT key
To store the new setting, display
end
and go to
step 6.
CAL key
To cancel the new setting and go to step 6.
Quitting the operation
6 The balance displays
Unit
. Press the CAL key to exit the programmable-unit function and return
to the weighing mode.
Using the function
Press the MODE key to select the programmable-unit (no display on the unit section). Perform
weighing as described in “5-1 Basic Operation (Gram Mode)”. After weighing, the balance displays the
result (weighing data in grams x coefficient).