11
Press "ADJ" to set the maximum value of 300K when the small yellow box is on the
left; press "ADJ" to clear when the small yellow box is on the right.
The minimum frequency is 50 Hz. When setting the frequency, the minimum bit is not
adjustable. When the frequency value is an integer multiple of the power frequency (50 /
60Hz) and the last two digits of the frequency display value are 0, the measured value is the
most accurate.
The frequency point can be set freely. Due to the limitation of hardware, it is
impossible for any point to be effective. When the invalid frequency point value occurs, the
system will adjust to the nearest frequency point of the set value. The higher the frequency,
the larger the step distance.
As shown in Figure 1-4, the intended frequency is 101700hz, but the actual frequency
is 101695 Hz as shown in Figure 1-5.
Figure 1-4 frequency setting 1
Содержание NJ300S
Страница 1: ...1 AAI 50Hz 300kHz Product manual NJ300S LCR impedance tester Accuracy Agility Instrument...
Страница 5: ...5 Operating temperature 5 40 Atmospheric pressure 860hPa 1060hPa...
Страница 6: ...6 Body description 9 1 2 3 4 5 7 8 10 11 6...
Страница 12: ...12 Figure 1 5 frequency setting 2 2 Calibration Figure 2 1 calibration page...
Страница 16: ...16 Figure 2 7 SMD fixture Figure 2 8 Confirmation prompt...