The above " Resistance measurement" in Figure 2-13 is the current measured
value of the equipment for "calibration resistance".
The following two parameter values of " Resistance cal value" in Figure 2-13
need to be manually filled in with the true value of this "calibration resistance".
(f) When the small yellow box is adjusted to the last bit of Xs, press "RIGHT"
again to start the calibration.
Figure 2-10 main test setting figure 2-11 main test page
Figure 2-12 calibration page figure 2-13 load calibration page
Note: the load calibration value can store 20 sets of data in total. When
the limit is exceeded, the previous value is overwritten.
Effective condition of calibration value:
a. The parameter setting (level, internal resistance, frequency) during
measurement must be completely consistent with the parameter setting during
calibration;
b. The impedance value of the tested part and the impedance value of the
"calibration resistance" used in the calibration shall not be too different.