17
(
4
)、"LOAD": Load calibration. When the user has higher requirements on
the measurement accuracy, the user can calibrate the measured value by himself.
After calibration, the measured value near the “calibration resistance” value
will be more accurate. (Here "calibrated resistance" means standard parts, such
as resistance, inductance, capacitance)
Figure 2-9 Load options
Calibration steps:
(a) Do open short circuit calibration.
(b) Clamp the standard resistor correctly.
(c) Enter the main test setting (Figure 2-10) and set the relevant parameters
(general default value).
(d) Press "OK" to enter the main measurement (Figure 2-11), and press "ADJFRQ"
to set the frequency. After the frequency setting is completed, the system
measures immediately.
(e) When the color of the main measurement parameter turns yellow, press
"BACK" to the main measurement setting page, then press this "NEXT" to switch
to the calibration surface (Figure 2-12), and finally enter the "LOAD" page
(Figure 2-13).