User open: Open circuit calibration data made by the user himself.
User short: Short circuit calibration data made by the user himself.
Sys short: the system has short circuit data.
Sys open: the open data of the Kelvin fixture.
Sys0603open: The open data when the opening distance of the patch fixture
is 0603 package length.
Sys0805open: Open data when the opening distance of the patch fixture is 0805
package length.
Sys1206open: Open data when the patch fixture opening distance is 1206 package
length.
Figure 2-6 shows the appearance of the Kelvin fixture. Figure 2-7 is the patch
fixture, the distance between the two needle tip openings at "A" in the figure
is the package length. SMD fixtures have better open circuit data than Kelvin
fixtures for measurement results.
After selecting a certain item of data and pressing "OK", there will be a
prompt as shown in Figure 2-8.
Figure 2-6 Kelvin fixture