3M
™
Advanced Systems Tester 900AST Series
27
Special Resistance
The Special Resistance function measures whether
a pair of conductors are electrically equal in
resistance. Use this function to measure the loop
resistance between the Red and Green test leads, the
resistance of each conductor connected to the Red
and Green test leads, and the difference in resistance
between the two conductors.
Connect the Red lead to one conductor of the pair
and the Green lead to the other conductor. Connect
the Black lead to a reference wire. Note that the
“reference wire” can be a separate wire, ground, or
the shield of the cable.
Short the other end of the
two conductors to each other and to the reference
wire.
Press the
key to start the test.
After the 900AST tester displays the results, press
the
key to repeat the test. Special Resistance is
not a continuous test.
Use the
key to return to the Toolbox.