Load Test
This function measures the DC current flowing
through a simulated load lamp inside the 900AST
tester. Connect the Red lead to the conductor under
test and the Black lead to a return or ground. The
900AST tester will continuously display the current
passing through the load as well as a bar graph
representing the approximate level of load lamp
illumination.
If the current in the load is greater than 110 mA, the
following ‘Current Warning’ screen is displayed.
This screen indicates that a high current has been
detected between the test leads and that the 900AST
tester has opened an internal relay to protect itself
from damage. Use standard safety practices for
disconnecting the test leads. Press the
key to
restart the 900AST tester. Fix problems before
restarting the ‘Load’ measurement.