96
3D SYSTEMS, INC
.
PN:80-D27 Rev.B
6 . Move down the sintering piston about 5-10mm .
7 . Open the airlock .
8 . Open the airlock door .
9 . Load the calibration plate on the sintering plate via the airlock .
10 . Close the airlock door .
11 . Close the airlock .
12 . Launch the inerting process up to reach the
fume safety
level .
13 .
Download the specific manufacturing FAB file on the printer.
14 . Start the draw (layer#0) of the calibration plate
with the red point
in order to control its positioning . Adjust the
calibration plate if necessary .
15 . Start the draw (layer#0) of the calibration plate
without the red point
.
16 . Repeat the previous step, to be sure the draw will be well marked .
17 . When job is complete, repeat steps from 6 to 8 to remove the calibration plate from the printer via the airlock .
18 . Close the airlock door and airlock . Then stop the inerting process .
19 . Inspect visually the sintered lines are homogeneously well marked over the whole surface of the plate (1) . If the result is
not correct (2), repeat the draw on a new calibration plate . If the result is still no correct, contact 3D Systems Customer
Support .
20 . The control of the calibration requires the current
correction file. To export this correction file:
•
Access to the
Configuration
tool,
•
Go to the
Optic
menu (1),
•
Click on the
Export
button (2),
•
Select "Correction file" from the file list, then save
the file on a folder from your network or a USB
key .
2
1
Calibration plate with crossing-lines homogeneously well marked
1
Calibration plate with crossing-lines
not marked correctly
2