4.4

 

SYSTEM CHECK 

The  SYSCHK  program  checks  the  performance  of  the  analog  cards,  as  well  as  the  overall 
operation of the GDP-32

24

.  It confirms analog card consistency while acquiring data, as well as 

any performance drift since the last calibration.  This check should be performed before and after 
any  field  job  in  order  to  ensure  that  the  data  collected  is  of  the  highest  quality.    To  start  the 

system check program press 

 from the Zonge Logo screen.   

The system check can be performed with various parameter settings in order to run either a quick 
check, or a more stringent check of the system.  For a quick check, the following parameters can 
be set.  These settings should provide a quick Go-NoGo run without the need to closely inspect 
the  overall  data.    Any  suspected  error  will  halt  the  automatic  progression  through  the 
frequencies,  of  the  AUTO  METHOD,  and  signal  the  operator  with  an  audible  warning.    The 
system should then be investigated more closely to confirm and identify and real problem.  See 
the System Check section of this manual for more information.   

FREQ

 

Set to the lowest frequency to be used for the job. 

CYCLES

 

At least 2 cycles for the lowest frequency. 

STACKS

 

At least 2, in order to quickly check analog card repeatability. 

AGAIN

 

Set to MAN for speed. 

NOTCH

 

Set  to  OUT.    Notch  filters  can  cause  wide  variability  around  harmonic 
frequencies of the filter.  Setting filters OUT, will minimized the chance of the 
automatic  sequencing  stopping  for  false  errors,  allowing  for  lower  detection 
limits to be set. 

VOLTS

 

Set  to  0.01 v. A low voltage  such  as  this can  help magnify  any  errors  in the 
analog cards. 

GROUP

 

This  will  be  set  automatically  for  this  purpose.    Stacks  of  data  will  be 
averaged  by  group  for  comparison  with  the  Data  Viewer.    Different  Groups 
collected  at  different  times  can  then  be  compared  to  find  any  change  in  the 
system. 

METHOD

 

Set to AUTO. 

dθ LIM

 

Set  to  0.1.   Without  NOTCH  filters, this  should provide  a good check. With 
NOTCH filters a setting of 1 is appropriate. 

dCal% L

 

Set to 1.  Without NOTCH filters  

After the settings are made, press 

 to proceed with the test.   When finished, press 

 

to return to the Zonge Logo screen.   

Enter

Summary of Contents for GDP-32

Page 1: ...ANUAL SYNCHRONIZATION 10 4 8 SET TIME 11 4 9 SYNC CHK 13 4 10 FIELD SURVEY PROGRAM 14 4 11 JOB_INFO SCREEN 15 4 12 LINE SETUP 16 4 13 ACQUISITION CONFIGURATION 17 4 14 CHANNEL TABLE 18 4 15 CHECK FIEL...

Page 2: ...ck Field connections Acquire Data Check Data 6 Transfer Data Simple menus make operating the GDP 3224 straightforward The easiest way to become familiar with the receiver operation is to try the vario...

Page 3: ...r 2015 CURSOR POSITION To position the cursor on the LCD display use the Cursor Control Keys or and To cycle through available options labels and numbers use the Rotary Entry Keys and Pg Dn NEXT FIELD...

Page 4: ...ith an XMT transmitter controller both the receiver and XMT should be powered on for at least an hour before syncing and acquiring data Press the button located on the Front Panel The unit beeps and t...

Page 5: ...sure all the channels to be calibrated are turned on 3 Set the frequency to the lowest frequency that will be used during the survey 4 Select the notch filter to be used for the survey Note It is good...

Page 6: ...ecially as the environment changes will provide the best results Refer to Section 6 Setup and the sections for specific programs for more information about calibration and when to use METHOD 1SHOTCAL...

Page 7: ...closely to confirm and identify and real problem See the System Check section of this manual for more information FREQ Set to the lowest frequency to be used for the job CYCLES At least 2 cycles for t...

Page 8: ...GDP 3224 INSTRUCTION MANUAL September 2015 Section 4 Page 8 GDP 3224 0404a_24SYSCHK_Quick For more detailed instructions see Section 8 System Check...

Page 9: ...zation the receiver should be connected to either a GPS antenna or to an XMT transmitter See Section 6 Receiver Setup for more information about making the connections to a transmitter controller 0405...

Page 10: ...the receiver and XMT should be powered on for at least an hour before running SYNC If a fast beep sounds while pressing any of the crystal adjust keys no more adjustment is available on that particul...

Page 11: ...the and keys to make adjustments After making adjustments move the cursor back to the MODE field and make sure the GPS Local Time appears correct If not adjust the Local Time Differential setting When...

Page 12: ...GDP 3224 INSTRUCTION MANUAL September 2015 Section 4 Page 12 GDP 3224 0408b_24SYNC_TimeManl...

Page 13: ...the internal reference Set channel 2 comp field to XMT and connect the channel inputs to the XMT Sync Out ports Turn Off all other channels not connected to a synchronized signal Press 0409a_24SNCCHK...

Page 14: ...eters can be entered directly on the Main Menu screen Various submenu screens are accessed directly from the Main menu for entering more specific job or special setup parameters Functions such as Cali...

Page 15: ...job information submenu where Operator Transmitter and other job based data are entered Programs such as TEM also use this screen to enter loop antenna configurations delay times and other data which...

Page 16: ...all programs contain all the fields listed below some programs contain additional fields but all the fields in this portion of the screen should be inspected and modified if necessary TIME SERIES Set...

Page 17: ...if necessary FREQ Select the frequency to analyze CYCL Set the number of cycles to average STACKS Set the number of repetitions of data acquisition to perform Each stack will be recorded as its own b...

Page 18: ...l parameters Some examples are Note Refer to the appropriate Survey Program chapters for more information on setting up channels Any designator other than OFF will turn on channels for all programs wi...

Page 19: ...ELD CONNECTIONS The Channel Table is also used to verify electrode and coil field connections prior to data acquisition Pressing starts the Contact Resistance CRES function to measure the resistance o...

Page 20: ...to evaluate the process When data has been acquired the final results will be calculated and displayed When taking a single stack of data the GDP will display the message Press ESCAPE to discard CONT...

Page 21: ...mode Each program has its own cache viewing features and procedures Refer to the appropriate field survey manual for more information Press the function key to return to Data Acquisition mode 0417a_2...

Page 22: ...screen press to start the connection 0418a_GDPNET_ManualStart Enter an IP address consistent with the address of the PC or laptop The first 3 octets for the receiver address should be set to match tho...

Page 23: ...er to the Local Area Network hub with a standard cable From the Zonge Logo screen press to start the connection Press to select the AUTOMATIC settings 0418c_GDPNET_NetStart Press to initiate the conne...

Page 24: ...attributes and properties from the PC or laptop computer s File Explorer This can make the GDP system unstable File attributes on the GDP must NEVER be set to Read Only or Hidden FILE TRANSFER Files c...

Reviews: