OPERATION
Carl Zeiss
Core accessories
Stemi DV4 / DR
3-16
B 40-003 e 08/00
•
Connect supply and signal cables, if required. If the camera will be used for a long period of time, it is
recommended to use the line adapter which is enclosed with the camera!
•
Switch on the camera.
•
Set the camera lens to
∞
via the distance setting control. Always check the "
∞
" setting before taking
exposures!
•
Zoom the camera lens from "W" to "T" until the maximum telelens focal length of 21 mm is
achieved. The image is trimless in this position.
☞
When the camera is switched off and the instrument is in the "PLAY" mode, the camera lens
moves back into the camera. After switching on STILL and MOVIE again, the lens is in the
wide-angle range. Therefore, the maximum telelens focal length must then be set again.
•
The camera is ready for operation.
☞
When taking photos in transmitted light with a lamp voltage close to U
max
(blue LED of the
illumination control is blinking!), remember that the lamp voltage is reduced to constant
voltage after three minutes!
•
For the operation of the camera, see the operating manual "SONY Digital Still Camera Cybershot
DSC-S70"!
☞
If other cameras with objective/filter thread M37 or M52 shall be attached via the eyepiece
adapter M37 / M52, you must check whether the images are format-filling without trimming
or vignetting, i.e. whether position and size of the exit pupil of the eyepiece adapters match
the focal length-dependent entry pupil!
This is checked as follows:
−
Vary the distance between the camera lens and the lens mount in steps, i.e. move the
sliding mount with camera on the lens mount in the eyepiece adapter by defined steps.
−
Zoom camera lens from wide-angle position (W) through to tele position (T).
−
Perform test until image is format-filling without trimming or vignetting.
If no format-filling image without trimming or vignetting is obtained, adaptation of the other
camera is not possible!
Summary of Contents for Stemi DR 1040
Page 1: ...Stereomicroscopes Stemi DV4 Stemi DV4 SPOT Stemi DR 1040 Stemi DR 1663 Operating Instructions...
Page 6: ...ILLUSTRATIONS Carl Zeiss Stemi DV4 DR 0 6 B 40 003 e 08 00...
Page 10: ...INTRODUCTION Carl Zeiss Stemi DV4 DR 1 4 B 40 003 e 08 00...
Page 18: ...DESCRIPTION Carl Zeiss Stemi DV4 DR 2 8 B 40 003 e 08 00...