Raise  /lower  the  condenser  apparatus  by  turning 

the  small  black  knob  (3c

  -  on  the  left  back 

side of the condenser below the stage) until the circumference of the light image is "sharp and 
well defined as a hexagon shape." 

 

* Note:

 During this process, the particular condenser ring optics can be selected to match the 

type of imaging desired: 

turn

 

the condenser (turret) control ring (3d)

 to: 

HDIC

 for brightfield 

work; 

DIC

 for differential interference contrast work.  For Phase Contrast work, use no “

1

” for 

10X, no “

2

” for 20X and 40X and no “

3

” for 63X objective. 

 

**Note:

 it might not be necessary to adjust/focus the condenser for different objective lenses 

(magnifications).  For  example,  if  condenser  is  focused  for  20x,  that  could  be  "acceptable"  for 
10x and 40x (air objectives); focus for 63x could be "acceptable" for 100x (oil objectives), but 
NOT for the 10, 20, and 40x air. Minor re-centering with the condenser centering screws can be 
performed: 

the light is best centered

 when the spot is opened up with the field diaphragm ring 

to 

about  80-  90%  full  field  illumination

,  and  the  light  circumference  is  "concentric"  with  the 

edge of the field boundary.  

***Note

:  the  silver  ring  on  the  condenser  opens  (clockwise)  and  closes  (counter-clockwise)  a 

diaphragm  for  more  or  less  light,  respectively  —  lesser  light  might  give  best  image  (totally 
counter-clockwise), however, adjust as necessary for the conditions.  

 

Before

 

After

 

Sharp edge

 

DIC

 

Brightfield

 

Phase 1

 

Phase 3

 

Summary of Contents for Axioskop

Page 1: ...ZEISS AXIOSKOP Microscope s User Manual Managed by For information about this instrument please contact Dr Alloysius Budi Utama ext 8232 or e mail budiutama rice edu...

Page 2: ...ed next to switch 1b The halogen lamp is in the vented box at the bottom rear of the scope 1c 2 For fluorescence microscopy work press green on off light switch 2a on the front of the arc lamp power s...

Page 3: ...he field diaphragm so they are not in the light path Note for Brightfield DIC adjustment the black horizontal metal filter slider should be in the no filter open position last position on left or 5th...

Page 4: ...different objective lenses magnifications For example if condenser is focused for 20x that could be acceptable for 10x and 40x air objectives focus for 63x could be acceptable for 100x oil objectives...

Page 5: ...en light color temperature of 3200 K into natural outdoor light color temperature of 5500 K When push buttons are IN or OUT the corresponding filters are IN or OUT as well Fluorescence microscopy Afte...

Page 6: ...orescence that contain various excitation filter dichroic mirror emission filter sets from right to left 450 490 um ex green em 510 560 nm ex red em 450 490 nm ex green yellow em 300 390 nm ex blue em...

Page 7: ...ne AF594 Texas red type fluorophores Excitation filter 510 560 nm Dichroic mirror 580 nm Emission filter 590 nm LP Third position for YFP type fluorophores Excitation filter 450 490 nm Dichroic mirror...

Page 8: ...contrast condenser position DIC Plan NEOFLUAR 40X NA 0 75 Ph 2 air n 1 Phase contrast condenser position 2 Brightfield condenser position HDIC DIC contrast condenser position DIC Plan NEOFLUAR 63X NA...