background image

When  the  arc  lamp  is  turned  on,  and  the  slide/specimen  is  placed  in  position  on  the  stage, 
illumination of the sample from the objective lens will be readily apparent. The slider, black and 
marked L on left side and R on right side, can be used to block / interrupt the illumination beam 
path (move to far right position) when not viewing the sample to prevent photobleaching. The 
green filter/middle position slightly filters the light, but the open position (far right) is probably 
best when examining the sample.  

*Note:

 in fluorescence microscopy, the less exposure of the sample / fluorophore to excitation 

light, the less photobleaching and subsequent loss of fluorescence and intensity of signal.  

The fluorescence reflector or tiller slider [large horizontal black metal box between ocular and 
objective lenses] has 4 positions for fluorescence that contain 

various excitation filter / dichroic 

mirror / emission

 

filter sets

, from right to left:  

 

 
 
 

// ~450-490 um ex. 

(green em.)

 

// ~510-560 nm ex

.  (red em.) 

 

// ~450-490 nm ex.  

(green/yellow em) 

 

// ~300-390 nm ex.  

(blue em.) 

 

 
Decide  which  filter  set  is  appropriate/best  for  exciting  and 
detecting the fluorophore(s) being used, and move the filter slider 

to that position. [See fluorescence reflector slider sheet for filter set details.]  
 

*Note:

  It  should  not  be  necessary  to  change/adjust  the  illumination  beam,  however,  the 

following  controls  affect  the  diameter  and  centering  of  the  light.  Levers  for  adjusting  the 

luminous field diaphragm

 (upper, left and right side) — move by pushing or pulling levers 

to 

the right to reduce and to the left to increase

 the illuminated area [“normally," the area of light 

should just fill the field of view, as for the halogen lamp beam with light microscopy]. Use 

the 

two silver centering screws 

(lower, left and right side) to position the luminous field diaphragm 

- to move and align the light beam, alternate twisting both screws. 

 

 

Summary of Contents for Axioskop

Page 1: ...ZEISS AXIOSKOP Microscope s User Manual Managed by For information about this instrument please contact Dr Alloysius Budi Utama ext 8232 or e mail budiutama rice edu...

Page 2: ...ed next to switch 1b The halogen lamp is in the vented box at the bottom rear of the scope 1c 2 For fluorescence microscopy work press green on off light switch 2a on the front of the arc lamp power s...

Page 3: ...he field diaphragm so they are not in the light path Note for Brightfield DIC adjustment the black horizontal metal filter slider should be in the no filter open position last position on left or 5th...

Page 4: ...different objective lenses magnifications For example if condenser is focused for 20x that could be acceptable for 10x and 40x air objectives focus for 63x could be acceptable for 100x oil objectives...

Page 5: ...en light color temperature of 3200 K into natural outdoor light color temperature of 5500 K When push buttons are IN or OUT the corresponding filters are IN or OUT as well Fluorescence microscopy Afte...

Page 6: ...orescence that contain various excitation filter dichroic mirror emission filter sets from right to left 450 490 um ex green em 510 560 nm ex red em 450 490 nm ex green yellow em 300 390 nm ex blue em...

Page 7: ...ne AF594 Texas red type fluorophores Excitation filter 510 560 nm Dichroic mirror 580 nm Emission filter 590 nm LP Third position for YFP type fluorophores Excitation filter 450 490 nm Dichroic mirror...

Page 8: ...contrast condenser position DIC Plan NEOFLUAR 40X NA 0 75 Ph 2 air n 1 Phase contrast condenser position 2 Brightfield condenser position HDIC DIC contrast condenser position DIC Plan NEOFLUAR 63X NA...

Reviews: