Use of the Axioskop (Zeiss)

 

The  Axioskop  is  equipped  for  viewing  samples  with 

transmitted  light  (from  a  halogen  lamp 

source)

  and  has  optics  for: 

Bright  field,  Phase  Contrast  and  Differential  Interference  Contrast 

(DIC) 

microscopy. Samples can also be viewed for 

fluorescence microscopy

 

using the incident-

light system (front a mercury lamp source).

  

1)

 

For transmitted light microscopy work

, press the 

green on/off light switch (1a) at the back-right side of the 
microscope  while  light  intensity  /  level  knob  is  located 
next  to  switch  (1b).  The  halogen  lamp  is  in  the  vented 
box at the bottom rear of the scope (1c). 

2) 

For fluorescence microscopy work

, press green on/off light switch  (2a) on the front of the 

arc lamp power supply box behind the scope (right side- 2b). The mercury lamp is in the vented 
housing on the rear of the scope and is above the halogen lamp box (2c).  
*Note: the halogen lamp does not need to be on when only viewing fluorescence 

 

 

Transmitted light microscopy (Brightfield/DIC) 

Begin  with  glass  slide/specimen  securely  in  place  on  stage,  light  at  medium 

brightness/temperature (~6).  
Ocular's (eyepieces) diopter rings set 
at "white dot-to-0-position". 

In most 

cases start with 10 x objective lens in 

2a

 

2b

 

2b

 

2c

 

1a

 

1b

 

1c

 

Summary of Contents for Axioskop

Page 1: ...ZEISS AXIOSKOP Microscope s User Manual Managed by For information about this instrument please contact Dr Alloysius Budi Utama ext 8232 or e mail budiutama rice edu...

Page 2: ...ed next to switch 1b The halogen lamp is in the vented box at the bottom rear of the scope 1c 2 For fluorescence microscopy work press green on off light switch 2a on the front of the arc lamp power s...

Page 3: ...he field diaphragm so they are not in the light path Note for Brightfield DIC adjustment the black horizontal metal filter slider should be in the no filter open position last position on left or 5th...

Page 4: ...different objective lenses magnifications For example if condenser is focused for 20x that could be acceptable for 10x and 40x air objectives focus for 63x could be acceptable for 100x oil objectives...

Page 5: ...en light color temperature of 3200 K into natural outdoor light color temperature of 5500 K When push buttons are IN or OUT the corresponding filters are IN or OUT as well Fluorescence microscopy Afte...

Page 6: ...orescence that contain various excitation filter dichroic mirror emission filter sets from right to left 450 490 um ex green em 510 560 nm ex red em 450 490 nm ex green yellow em 300 390 nm ex blue em...

Page 7: ...ne AF594 Texas red type fluorophores Excitation filter 510 560 nm Dichroic mirror 580 nm Emission filter 590 nm LP Third position for YFP type fluorophores Excitation filter 450 490 nm Dichroic mirror...

Page 8: ...contrast condenser position DIC Plan NEOFLUAR 40X NA 0 75 Ph 2 air n 1 Phase contrast condenser position 2 Brightfield condenser position HDIC DIC contrast condenser position DIC Plan NEOFLUAR 63X NA...

Reviews: