START-UP
Axioskop 2
mot plus
Attachments and conversions
Carl Zeiss
B 40-075 e 10/03
2-21
•
Loosen the fixing screw (2-33/
5
) on the heat
6
) and pull off the heat sink. If the
heat sink does not come off (or cannot be
replaced), unscrew the fixing screw completely,
screw into the next hole and press down the
heat sink with it.
•
Dispose of the defective lamp properly.
•
Insert the new lamp of smaller diameter into
the heat sink as far as the stop.
Pay attention to the correct hole
diameter when inserting and do not
apply force to the lamp.
The holder in the heat sink is marked
with
H
(for HBO) or
X
(for XBO).
•
Clamp the fixing screw on the heat sink.
•
1
) with heat sink
4
) pressed down in
the holder provided in the illuminator socket
(2-33/
). Make sure that the power cable to the
heat sink is free at the side. Release the spring
clip. Do not apply force to the lamp.
•
Fit the illuminator housing (2-32/
) on the lamp
3
) and clamp with the clamping
).
•
Connect HBO 103 plug and line plug to the
HBO 100 W transformer.
•
Adjust HBO 103 illuminator as described in
section 2.1.11 (3).
Fig. 2-32
Removing the housing of the
HBO 103
Fig. 2-33
Changing the HBO 103 W/2
mercury pressure short-arc lamp