Axioskop 2
plus
START-UP
Axioskop 2
mot plus
Attachments and conversions
Carl Zeiss
B 40-075 e 10/03
2-15
2.2
Attachments and conversions
☞
The steps required for attachments to and conversions of the microscope are explained in the
following. After this, the respective modules must be reset in a functioning status again.
2.2.1
Changing the condenser
•
Move stage carrier to the upper stop position
via the focusing drive.
•
8
) to fold out the front lens on
the available condenser or, in the case of the
motorized version, swing it out by pressing the
respective key.
•
) to the lowest
position via drive for height adjustment
).
•
In the case of the motorized condenser
(Axioskop 2
mot plus
), loosen screwing ring
, on the left of the stand) of connecting
cable and pull off the connecting cable.
•
Unscrew both centering screws (2-22/
3
) until
condenser can be easily taken out of its mount.
•
Lift available condenser from the orientation
groove by slightly tilting it upwards and pull it
out to the front.
•
) to fold out the front lens on the condenser to be inserted or, in the case of the
motorized version, swing it out by pressing the respective key.
•
1
) between condenser carrier (2-22/
2
4
). Orient the
locking screw on the underside of the condenser in the direction of the groove (2-22/
7
).
•
Press dovetail of condenser against the spring mount (2-22/
) of the condenser carrier until the
condenser can be placed on the condenser carrier horizontally.
•
Let condenser go smoothly; the screw will lock in position in the groove (2-22/
) at the front.
•
Screw in centering screws until they engage in the dovetail.
Fig. 2-22
Changing the condenser