Axioskop 2
plus
CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axioskop 2
mot plus
Instrument maintenance
Carl Zeiss
B 40-075 e 10/03
4-3
4.2 Instrument
maintenance
4.2.1 Performing
checks
•
Make sure that the instrument(s) comply with the required line voltage (e.g. the HBO 100
transformer).
•
Check line cable and plug for defects.
•
If any damage is visible, switch off the instrument. Have instrument repaired only by an expert.
•
Make sure that the maximum operation time of the halogen and mercury vapor short-arc lamps is not
exceeded.
4.2.2
Changing the fuses
Always pull the power plug before
changing the fuses.
The fuse compartment is positioned at the
microscope rear. It is combined with the
instrument socket and contains two
T4 A/H
250 V
fuses.
•
Disconnect the instrument from the line.
•
Pull out the fuse holder (4-1/
1
) from the front.
Use a small screwdriver if necessary.
•
Remove fuses from the holder and insert new
fuses.
•
Push the fuse holder into the fuse compartment
(4-1/
2
) to the stop.
•
Connect the instrument to the line.
Fig. 4-1
Changing the fuses