For confirmation that disinfection has been carried out, it is advised to
place a visible sign on the running surface reading
"disinfected"
.
Fluids required for disinfecting and cleaning must be stored, prepared,
and kept ready for use exclusively in the containers provided, in order
to avoid them being mistaken for other fluids.
On no account bring any disinfection fluids or other liquids in direct
contact with the FDM-T sensor when cleaning. Should any liquid enter
the platform it is likely to be damaged irreparably.
If you apply disinfectant be sure to strictly follow the recommendations
provided by the manufacturer of the disinfectant. Especially consider the
rules concerning the disinfectant’s recommended duration of impact.
Please make sure to switch off the device and pull the mains plug out
of the socket before you commence disinfecting and cleaning.
Do not use any aggressive agents to clean the measuring system.
Seite 40/69
Seite 40/41
FDM-T-Sports
6.5. Cleaning and disinfection
6.5.1. Cleaning procedure
The treadmill and accessories are cleaned with a moist cloth while the device is switched off
and the mains plug taken out.
6.5.2. Disinfection procedure
The treadmill can be disinfected by wiping over with suitable detergents. Best wipe the running belt
or other parts of the treadmill with a cloth soaked with disinfection liquid. To remove more resistant
contamination directly spray the running belt with disinfection liquid.
Recommended disinfection agent:
Composition approx. 25% ethanol, 35% propanol
e.g., Mikrozid Liquid/Schülke & Mayr or similar solutions