All FDM-T systems can be stored without power supply for a maximum
of 6 to 9 months. After this period, the battery may be totally discharged
due to lacking power supply. If the storage of the device exceeds this period,
a reprogramming of the treadmill control may be necessary.
FDM-T systems must NOT be operated in wet zones, wet rooms
(swimming pools, saunas) or climatic chambers.
Direct contact with liquids must always be avoided, as the measuring
system is not protected against the entry of liquids. Liquids penetrating
the device can cause fire, electrical shock, or other severe accidents.
The FDM-T system is NOT specified for the operation in vacuum, hyper-
baric or altitude chambers.
The measuring systems are not intended for operation in potentially
explosive atmospheres of medically used rooms or oxygen-enriched
atmospheres.
The devices must not be operated in proximity to e.g., engines or
transformers with a high connected load as well as mains current
lines, as electrical or magnetic interference fields can falsify correct
measurements resp. turn them impossible.
Therefore, the devices must be protected against humidity.
The ventilation slots of the treadmills must be always free, so that
air can circulate freely.
Seite 7/69
Seite 7/41
FDM-T-Sports
2 Safety
2.1. Environmental conditions
FDM-T measuring systems are suitable for application in dry interiors
with level ground such as those in hospitals, doctors’ surgeries, and laboratories.
Temperature
10°C to 40°C
Relative humidity
30% to 70%
Air pressure
700 hPa to 1100 hPa
2.2. Storage and transport
Storage and transport of the measuring system are only to be affected in the original
packaging provided by zebris.
Temperature
-20°C to +70°C
Relative humidity max. 95%, non-condensing
Protect from moisture