6.2. Mandatory periodic inspections and STK
■
For maintaining the correct state of the electrical equipment, checks and technical
safety inspections must be carried out repeatedly (e.g., within Germany, acc. to
BGV A3, and accident prevention regulations and technical safety tests according
to the Medical Device Operating Regulations). Here it should be noted that standard regulations for
electrical devices are concerned here and not measures that are specific to zebris.
■
Before each use of the measuring system, it is recommended for safety reasons to check the correct
state of all the connection cables, as well as the mains cable, mains plug and mains socket. Should
certain parts be damaged, they must be replaced before continuing to use the measuring system.
■
If the name plate or other important labels (e.g., warning notes) are damaged or illegible they have
to be replaced by the manufacturer for safety reasons.
■
Each treadmill has a slip-resistant surface on both sides next to the running surface.
These treads provide secure footing when descending in an emergency. In particular, the surface
must be kept dry and dried if necessary. When worn, the treads must be replaced.