44
APPENDIX D CALIBRATION TABLES
Calibration values for various atmospheric pressures and altitudes.
PRESSURE
ALTITUDE
CALIBRATION
inches Hg
mm Hg
kPa
Feet
m
VALUE
30.23
768
102.3
-276
-84
101
29.92
760
101.3
0
0
100
29.61
752
100.3
278
85
99
29.33
745
99.3
558
170
98
29.02
737
98.3
841
256
97
28.74
730
97.3
1126
343
96
28.43
722
96.3
1413
431
95
28.11
714
95.2
1703
519
94
27.83
707
94.2
1995
608
93
27.52
699
93.2
2290
698
92
27.24
692
92.2
2587
789
91
26.93
684
91.2
2887
880
90
26.61
676
90.2
3190
972
89
26.34
669
89.2
3496
1066
88
26.02
661
88.2
3804
1160
87
25.75
654
87.1
4115
1254
86
25.43
646
86.1
4430
1350
85
25.12
638
85.1
4747
1447
84
24.84
631
84.1
5067
1544
83
24.53
623
83.1
5391
1643
82
24.25
616
82.1
5717
1743
81
23.94
608
81.1
6047
1843
80
23.62
600
80.0
6381
1945
79
23.35
593
79.0
6717
2047
78
23.03
585
78.0
7058
2151
77
22.76
578
77.0
7401
2256
76
22.44
570
76.0
7749
2362
75
22.13
562
75.0
8100
2469
74
21.85
555
74.0
8455
2577
73
21.54
547
73.0
8815
2687
72
21.26
540
71.9
9178
2797
71
20.94
532
70.9
9545
2909
70
20.63
524
69.9
9917
3023
69
20.35
517
68.9
10293
3137
68
20.04
509
67.9
10673
3253
67
19.76
502
66.9
11058
3371
66
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...