System Setup
Section 6
YSI, Incorporated
Model 52
21
Press CONFIRM to select communications setup. The following display will appear:
Use SKIP to go on to the next choice, or press CONFIRM to change the baud rate. If
you pressed CONFIRM, you would see:
Use
!
to increase the baud rate and
"
to decrease it. The Model 52 supports 110, 300,
1200, 2400, 4800, and 9600 baud and defaults to 9600. When you have selected the
correct baud rate, press CONFIRM to enter the value. The following display will appear:
Proceed exactly as you did for baud rate to set word length (7 or 8), then parity (odd, even
or none), then stop bits (0, 1, or 2).
6.4
LAST DIGIT SUPPRESSION
When measuring dissolved oxygen in the O
2
-TEMP function the SKIP key becomes a
toggle switch for suppressing last digit of the reading in both % and mg/L. When the last
digit is suppressed, the measurement will still be as accurate as it is when the last digit is
displayed.
6.5
CONTRAST ADJUSTMENT
The Model 52 has a contrast adjustment feature. This feature allows you to customize the
Model 52 display for comfortable viewing. To adjust the contrast set the function switch
to SYSTEM SETUP and press the SKIP key repeatedly until the following display
appears:
S e t u p
b a u d r a t e ?
S e t u p
b a u d r a t e = 9 6 0 0
S e t u p
w o r d l e n g t h ?
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...