System Setup
Section 6
YSI, Incorporated
Model 52
20
Use the arrow keys to select the desired autostable % (the smaller the %, the more
accurate your reading and the more time it will take to achieve). Press CONFIRM. The
following display will appear:
Press CONFIRM. The following display will appear:
Use the arrow keys to select the # of readings which must pass while complying with the
% change entered previously. (The larger the #, the greater the accuracy and the longer it
will take.)
6.3
SETTING COMMUNICATION PARAMETERS
The Model 52 will default to communications values that are appropriate for most
printers and personal computers. The defaults are Baud rate: 9600; parity = none; word
length = 8; and stop bits = 1. We recommend that you try these default settings before
attempting to change them. If you must change communication parameters on the Model
52, do so before using the RS232 port. Before beginning this sequence you must know
the baud rate, word length, number of stop bits, and parity of the PC in use. To find this
information, consult the documentation that accompanies your printer or PC. The Model
52 supports all common variations of these parameters.
To change a communication parameter, turn the function switch to SYSTEM SETUP.
Press SKIP repeatedly until the following display appears:
S e t u p
# o f r d g s ?
S e t u p
# o f r d g s = 5
S e t u p
c o m m u n i c a t i on ?
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...