![YSI Model 52 Operation Manual Download Page 20](http://html.mh-extra.com/html/ysi/model-52/model-52_operation-manual_915315020.webp)
Operation
Section 5
YSI, Incorporated
Model 52
16
5.6
PRINTING DATA FROM MEMORY
To send all data from the YSI Model 52 memory to a serial printer, attach the RS232
Cable (supplied) to the Model 52 and your printer. With the function switch set to O
2
-
TEMP press the
"
key repeatedly until a display similar to the following appears:
Press the CONFIRM key. All of the data currently stored in the Model 52 memory will
be sent to your printer.
Note: There is no way to send individual records from memory. All data from memory
will be sent in batch to the printer every time the “Print Data” command is confirmed.
5.7
SENDING RECORDED DATA TO A PERSONAL COMPUTER
To send data from the Model 52 memory to a personal computer (PC), run a
communications program such as ProComm™, CROSSTALK™, or PC Talk™, and
enter terminal mode. (See section IX). Connect the RS232 cable (supplied) to the Model
52 and your PC. With the function switch set to O
2
-TEMP press the
"
key repeatedly
until a display similar to the following appears:
Note: Be sure that the baud rate of the 52 and the communication program on your PC
are the same. (See section VIII.C.)
Press the CONFIRM key. All of the data currently stored in the Model 52 memory will
be sent to your PC.
Note: There is no way to send individual records from memory. All data from memory
will be sent in batch to the PC every time the “Print Data” command is confirmed.
7 . 0 m g / L 2 3 . 4 C
P r i n t D a t a ?
7 . 0 m g / L 2 3 . 4 C
P r i n t D a t a ?
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...