Calibration
Section 3
YSI, Incorporated
Model 52
10
Note: When you change the zero calibration point, you offset the other value, so that after
zeroing the probe for zero oxygen, you must recalibrate in an oxygen environment. If you
try to take readings without recalibrating, the display will provide this warning:
You may also select mg/L calibration and then zero-calibrate and the result will be the
same.
3.7
RESETTING THE DEFAULT ZERO CALIBRATION
To reset the zero calibration to its default value, turn the instrument off and then hold
down SKIP and
!
while turning the instrument back on again. The display below will
show briefly, then the POST operation and display will resume.
This procedure resets the meter to its original state. You must reset the time, date,
salinity, and the RS232 parameters after restoring the default zero. As with the zero
calibration, you must also recalibrate after restoring the default zero.
8 . 7 4 m g / L 2 5 . 6 C
R e c a l i b r a t e
S y s t e m r e s e t t o
d e f a u l t v a l u e s
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...