Salinity
Section 4
YSI, Incorporated
Model 52
12
Even if your sample is saline, it is still best to calibrate in air. However, if you must
calibrate in air-saturated water, it is best if the calibration sample is similar to the sample
you intend to measure. Disregard the mg/L reading when measuring non-aqueous liquids.
Such samples may have an oxygen solubility or Bunsen coefficient significantly different
from those used in the Model 52.
Note: If you are calibrating in mg/L, the salinity must be set before calibration.
If you are calibrating by a Winkler titration, the salinity of the calibration sample must be
the same as that of the samples you intend to measure. It is not necessary to know the
salinity value, only that it is the same in both the calibration and measured samples. You
must not change the salinity setting after calibration. If you do change it after a mg/L
calibration, you must recalibrate.
After calibrating, switch to O
2
-TEMP. The display will simultaneously show temperature,
% air saturation, and oxygen concentration.
6 . 3 3 m g / L 1 9 . 8 C
8 8 . 1 % 1 1 : 4 5
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...