Calibration
Section 3
YSI, Incorporated
Model 52
6
3.3
CALIBRATION IN AIR
Three separate calibration techniques are discussed here: calibration in air, calibration in
air-saturated water, and calibration by Winkler titration. Choose the one technique which
best fits your application. Calibration in air is the simplest and most accurate method of
calibration.
To calibrate in air:
1. Place a prepared probe in air at 100% relative humidity. To achieve this, BOD probes
can be placed in a BOD bottle with approximately 1” of water. Field probes can be placed
in the plastic calibration bottle with a moistened sponge, or calibrated with the 5075A
Calibration Chamber (see probe instructions Appendix A).
2. Switch to CALIBRATE. The display will read:
Select percent calibration by pressing CONFIRM. A display similar to the one below
will be shown.
3. Using
!
and
"
, adjust the number to the calibration value determined from the
altitude or pressure chart on the back of the Model 52. SKIP sets the value to 100.0%;
and shows the following display:
you may wish to do this to bring the setting quickly to a value near the one you want, then
continue with
!
and
"
. Pressing SKIP again sets the value to 0.0% (see Zeroing the
Probe). Pressing SKIP a third time brings it back to the last calibration value.
C a l i b r a t e
i n p e r c e n t ?
E n t e r c a l v a l u e
L a s t = 9 6. 4 %
E n t e r c a l v a l u e
S a t . = 1 0 0 %
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...