APPENDIX A GENERAL SPECIFICATIONS
Oxygen Measurement
Ranges:
0.0 to 19.99 mg/L dissolved oxygen
0.0 to 199.9% air saturation
Accuracy:
±0.1% of saturation value, plus 1 Least Significant Digit for mg/L
readings, plus probe error
±0.1% of air saturation, plus 1 Least Significant Digit for % readings, plus
probe error
Resolution:
0.1% of the saturation value, expressed in mg/L, or 0.01 mg/L , whichever
is greater; 0.1% air saturation
Temperature Measurement
Range:
-5.0 to +45.0°C
Accuracy: ±0.1°C, plus probe error
Resolution: 0.1°C
Temperature Compensation
The mg/L mode is automatically temperature-compensated to an accuracy of ±1% of DO
readings between 0 and 5°C, and to an accuracy of ±0.6% of readings between 5 and
45°C.
The % air saturation mode is automatically temperature-compensated to an accuracy of
±0.5% of calibration values between 0 and 5°C, and to an accuracy of ±0.3% of values
between 5 and 45°C.
Salinity Compensation
Range:
0.0 to 40.0 ppt
Accuracy: ±.02
mg/L
Operating Environment
0 to 45°C, 10 to 90% humidity, non-condensing
Water Resistance
With the probe receptacle capped, every case opening is has a gasket to resist the entry of
water.
Summary of Contents for Model 52
Page 1: ...YSI Model 52 Dissolved Oxygen Meter Operations Manual ...
Page 2: ......
Page 8: ...Preparing the Meter Section 2 YSI Incorporated Model 52 4 ...
Page 27: ...System Setup Section 6 YSI Incorporated Model 52 23 ...
Page 29: ...Personal Computer Interface Section 7 YSI Incorporated Model 52 25 ...
Page 35: ...Discussion of Measurement Errors Section 9 YSI Incorporated Model 52 31 ...
Page 41: ...Warranty and Repair Section 10 YSI Incorporated Model 52 37 ...
Page 45: ...Required Notice Appendix B YSI Incorporated Model 52 41 ...
Page 47: ...Oxygen Solubility Appendix C YSI Incorporated Model 52 43 ...
Page 49: ...Calibration Tables Appendix D YSI Incorporated Model 52 45 ...
Page 50: ...Calibration Tables Appendix D YSI Incorporated Model 52 46 ...