Principles of Operation
Section 8
or less have larger electrodes that are closely-spaced.
CHOOSING A CONDUCTIVITY CELL
Decide which cell will be the most useful for your conditions by considering the conductivity of
the solution you want to measure and the size of the sample.
0 0.1 1.0
10 100 1000
µ
S/cm
1 10
100
1000
mS/cm
3440
3401
3418
3256,
3255, 3446
3254,
3253,
3252,
Conductivity
The chart above reflects general guidelines. Refer to cell specifications for details.
The multipoint calibration option on the YSI Model 3200 Conductivity Instrument gives greater
flexibility and performance when using one cell over a wide conductivity range. This feature
allows users to calibrate one cell with up to 5 different conductivity standard solutions. When
analyzing a sample, the instrument selects the calibration value that is closest in conductivity to
the sample.
3200
S
ERIES
C
ONDUCTIVITY
C
ELLS
Dip cells are generally used for routine conductivity measurements. The 3254 Fill Cell is
designed for small sample or high throughput work. It requires only 5 ml of sample and can
take measurements quickly from one sample to the next.
The 3255 Flow Cell is designed for in-line conductivity measurements, such as for
ultrapure water systems.
YSI Incorporated
Model 3200
43
Summary of Contents for 3200
Page 2: ......
Page 30: ...System Configuration and Operation Section 3 YSI Incorporated Model 3200 26 ...
Page 40: ...Diagnostics Section 5 YSI Incorporated Model 3200 36 ...
Page 44: ...Troubleshooting Section 7 YSI Incorporated Model 3200 40 ...
Page 48: ...Principles of Operation Section 8 3252 3253 3254 3255 3256 YSI Incorporated Model 3200 44 ...
Page 56: ...Principles of Operation Section 8 YSI Incorporated Model 3200 52 ...
Page 59: ...Warranty and Repair Section 9 YSI Incorporated Model 3200 55 ...
Page 70: ...Accessories and Replacement Parts Section 10 YSI Incorporated Model 3200 66 ...
Page 74: ...Required Notice Appendix B YSI Incorporated Model 3200 70 ...
Page 76: ...RS232 Connections Appendix C YSI Incorporated Model 3200 72 ...
Page 82: ...Conversion Factors Appendix E YSI Incorporated Model 3200 78 ...
Page 85: ...Glossary of Terms Appendix F YSI Incorporated Model 3200 81 ...