System Configuration and Operation
Section 3
[0] # Configuration number
CONFIG
[CONFIG 0] Title
0.0000uS offset
[Single]
Cell
Cal
[1.00]
K
[Linear] Comp
[1.91] %/°C Comp val
[0.68] T.D.S. mult
[25.0] °C Comp Temp
ENTER TO SEND CONFIGURATION
UP
DOWN
DIGIT
NEXT
Press the [UP] or [DOWN] soft key to select one of the six (0-5) stored configurations. Each
configuration contains the single cell constant, zero offset, multiple cell constants, total dissolved
solids (TDS) multiplier,
linear temperature compensation coefficient
and
non-linear temperature
compensation data.
To change the currently selected configuration, press [NEXT] to move the cursor to the field you
want to change. Then, use the [UP], [DOWN] and [DIGIT] soft keys to change the field, except
Offset. The Offset field can only be changed from the Offset menu. The following table shows
the choices for each field.
Example: To select the standard non-linear temperature compensation curve for ultrapure water,
press [NEXT] three times, to move the cursor to the Comp field, then, press the [UP]
or [DOWN] soft key to cycle through the selections until Pure is displayed. Press
[MODE] two times to exit to the Operation menu.
Field
Range / options
Description
Title
A - Z or 0 - 9
User defined name for this configuration.
Cell Cal
Single or Multi
Selects between a single fixed cell constant
and multiple cell constants.
Comp
Linear, N. Linear, Pure, Natural
or Off
Selects between a linear temperature
compensation coefficient, non-linear
temperature compensation curve, standard
non-linear temperature compensation curve
(pure or natural) or no temperature
compensation.
TDS mult
0.55 - 0.90
Factor used to calculate the total dissolved
solids from the conductivity.
uS offset
0.0000 - 0.0150
Zero offset. Used only when measuring
conductance below 10uS to offset cell lead
insulation resistance. This value can only be
changed from the Offset menu.
K
0.001 - 100.999
Single fixed cell constant value (K). This
value is only used when the Cell Cal field is
set to Single.
YSI Incorporated
Model 3200
13
Summary of Contents for 3200
Page 2: ......
Page 30: ...System Configuration and Operation Section 3 YSI Incorporated Model 3200 26 ...
Page 40: ...Diagnostics Section 5 YSI Incorporated Model 3200 36 ...
Page 44: ...Troubleshooting Section 7 YSI Incorporated Model 3200 40 ...
Page 48: ...Principles of Operation Section 8 3252 3253 3254 3255 3256 YSI Incorporated Model 3200 44 ...
Page 56: ...Principles of Operation Section 8 YSI Incorporated Model 3200 52 ...
Page 59: ...Warranty and Repair Section 9 YSI Incorporated Model 3200 55 ...
Page 70: ...Accessories and Replacement Parts Section 10 YSI Incorporated Model 3200 66 ...
Page 74: ...Required Notice Appendix B YSI Incorporated Model 3200 70 ...
Page 76: ...RS232 Connections Appendix C YSI Incorporated Model 3200 72 ...
Page 82: ...Conversion Factors Appendix E YSI Incorporated Model 3200 78 ...
Page 85: ...Glossary of Terms Appendix F YSI Incorporated Model 3200 81 ...