Data Storage and Retrieval
Section 4
Comma Delimited Format, send (recalled data)
0 ,3/20/97, 5:37:17PM, 1234uS/cm ,23.45C ,K=0.09,1.91%/C, 9999uS
1 ,3/20/97, 5:46:15PM, 1234uS/cm ,23.45C ,K=0.09,1.91%/C, 9999uS
2 ,3/20/97, 5:57:34PM, 1234uS/cm ,23.45C ,K=0.09,1.91%/C, 9999uS
The two fields on the right of the 3200 setup send display, below temperature, allow the user to
select the measurement parameters that will be sent. Each of the two fields, (last two fields of the
report) can be set to read one of the following parameters:
•
Conductance
•
Conductivity
•
Resistance
•
Resistivity
•
Salinity
•
Total dissolved solids
NOTE: When sending recalled data, the units cannot be selected from the Setup send menu,
since they were already selected when the readings were previously stored.
After you have selected the parameters, press [MODE] once, to return to the Send Setup menu, or
three times to return to Operation mode.
DATA LOGGING
The 3200 can be setup to automatically send results to a computer or serial printer via the RS232
port at a user selected time interval from 5 to 1800 seconds (30 minutes).
From the Data menu, press the [SND SETUP] soft key, then press the [SET SEN] soft key. The
following screen will be displayed.
[ ] Field delimiter
SETUP SEND
[ ] Space units
[*] Temperature
[*]
Date
[Conductance]
[*]
Time
[Conductivity]
[ ] Cell config
[ ] Compensation
[0] S interval
UP
DOWN
DIGIT
NEXT
Setting the S interval parameter to a value other than zero programs the 3200 to automatically
send the current readings via the RS232 serial port. This allows data to be logged to a computer
or serial printer. The S interval is the number of seconds between transmissions. For example, if
the S interval is set to 30, every 30 seconds the 3200 will send the current readings over the
RS232 port.
YSI Incorporated
Model 3200
33
Summary of Contents for 3200
Page 2: ......
Page 30: ...System Configuration and Operation Section 3 YSI Incorporated Model 3200 26 ...
Page 40: ...Diagnostics Section 5 YSI Incorporated Model 3200 36 ...
Page 44: ...Troubleshooting Section 7 YSI Incorporated Model 3200 40 ...
Page 48: ...Principles of Operation Section 8 3252 3253 3254 3255 3256 YSI Incorporated Model 3200 44 ...
Page 56: ...Principles of Operation Section 8 YSI Incorporated Model 3200 52 ...
Page 59: ...Warranty and Repair Section 9 YSI Incorporated Model 3200 55 ...
Page 70: ...Accessories and Replacement Parts Section 10 YSI Incorporated Model 3200 66 ...
Page 74: ...Required Notice Appendix B YSI Incorporated Model 3200 70 ...
Page 76: ...RS232 Connections Appendix C YSI Incorporated Model 3200 72 ...
Page 82: ...Conversion Factors Appendix E YSI Incorporated Model 3200 78 ...
Page 85: ...Glossary of Terms Appendix F YSI Incorporated Model 3200 81 ...