![YSI 3200 Operation Manual Download Page 36](http://html.mh-extra.com/html/ysi/3200/3200_operation-manual_915297036.webp)
Data Storage and Retrieval
Section 4
The baud rate range is 1200 to 19200. After you have selected the parameters, press [MODE]
once, to return to the Send Setup menu, or three times to return to Operation mode.
To select the report format and which data fields will be sent over the RS232 serial port when the
[SEND] soft key is pressed, press the [SET SEN] soft key. The following screen will be
displayed.
[ ] Field delimiter
SETUP SEND
[ ] Space units
[*] Temperature
[*]
Date
[Conductance]
[*]
Time
[Conductivity]
[ ] Cell config
[ ] Compensation
[0] S interval
UP
DOWN
DIGIT
NEXT
Each parameter is selected by placing an asterisk character between the brackets in front of it.
Press the [NEXT] soft key to select the parameter that you want to change. Then press the [UP] or
[DOWN] soft key to turn the asterisk character on or off.
You can select the report format that you wish to use by changing the field delimiter. The default
is a space [ ]. This provides a standard text output. You may also select a comma [,] by pressing
the [UP] or [DOWN] soft key.
When Space units is selected, by placing an asterisk character between the brackets in front of it
[*], a space will be inserted between the value and the units. This is useful when importing the
data to a spreadsheet.
The following are examples of each send format:
Space Delimited Format, send (current display)
1/23/97 12:34:56AM 0# 1.23%/°C 12.34C 12.3mS 1.23ppt
12/12/97 12:34:56AM 0# 1.23%/C 12.34C 1234uS/cm 0.12ppt
Comma Delimited Format, send (current display)
1/23/97, 12:34:56AM, 0# ,1.23%/°C, 12.34C , 12.3mS ,1.23ppt
12/12/97, 12:34:56AM, 0# ,1.23%/C, 12.34C , 1234uS/cm ,0.12ppt
Space Delimited Format, send (recalled data)
0 3/20/97 5:37:17PM 1234uS/cm 23.45C K=0.09 1.91%/C 9999uS
1 3/20/97 5:46:15PM 1234uS/cm 23.45C K=0.09 1.91%/C 9999uS
2 3/20/97 5:57:34PM 1234uS/cm 23.45C K=0.09 1.91%/C 9999uS
YSI Incorporated
Model 3200
32
Summary of Contents for 3200
Page 2: ......
Page 30: ...System Configuration and Operation Section 3 YSI Incorporated Model 3200 26 ...
Page 40: ...Diagnostics Section 5 YSI Incorporated Model 3200 36 ...
Page 44: ...Troubleshooting Section 7 YSI Incorporated Model 3200 40 ...
Page 48: ...Principles of Operation Section 8 3252 3253 3254 3255 3256 YSI Incorporated Model 3200 44 ...
Page 56: ...Principles of Operation Section 8 YSI Incorporated Model 3200 52 ...
Page 59: ...Warranty and Repair Section 9 YSI Incorporated Model 3200 55 ...
Page 70: ...Accessories and Replacement Parts Section 10 YSI Incorporated Model 3200 66 ...
Page 74: ...Required Notice Appendix B YSI Incorporated Model 3200 70 ...
Page 76: ...RS232 Connections Appendix C YSI Incorporated Model 3200 72 ...
Page 82: ...Conversion Factors Appendix E YSI Incorporated Model 3200 78 ...
Page 85: ...Glossary of Terms Appendix F YSI Incorporated Model 3200 81 ...