3-9
If you are setting up to measure galactose or hydrogen peroxide using one of the YSI Enzyme
Membrane Sensors, see Section 4 for more information.
* If you use units other than g/L for the linearity test, refer to Appendix A for conversion
values.
NOTE: If any reading is outside of the specified tolerance limits, recalibrate and repeat the
linearity test. If the reading is still out of tolerance, refer to Section 8, Troubleshooting.