5-15
Press [0] to return to RUN mode setup.
RUN mode setup: 1-SampleProtocol
2-AutoStandby 3-AutoCal 4-Monitor
You have now completed Run mode setup. Press [0] to return to Select setup menu level.
Select setup: 1-General 2-MeasParameter
3-RunMode 4-Report 5-PrntSetup 6-Default
Press [4] for Report. The display will read:
Select sample report format: Brief
1-None 2-Brief 3-Detail
Press [ENTER] to confirm sample report format and move to cal report format.
Select cal report format: Brief
1-None 2-Brief 3-Detail
Report Format refers to the information that will be printed each time the 2700 SELECT runs a
sample or a calibration. The default settings are "Brief".
Detail report formats are typically used when you want to study probe currents and temperature
changes with newly installed reagents or enzyme membranes. You may also elect to choose
Detail report for the first few calibrations or samples after a long idle period where no samples
or cals are run. If you suspect problems with the instrument, you should run Detail reports to
provide more information for troubleshooting.
See Appendix D for example report formats.
Press [0] to return to Select Setup.
Select setup: 1-General 2-MeasParameter
3-RunMode 4-Report 5-PrntSetup 6-Default
You have completed Report setup.
Press [5] for PrntSetup. The display will read:
Printing instrument setup... Please wait
The instrument will print the setup information you have entered. The information is an
itemized list of the key parameters that you have chosen. The printing will take about 60
seconds and use about 8 inches (20 cm) of printer paper. See Appendix C for a sample printout.
When printing is completed, the instrument returns to Select setup menu.