IM 12D06A03-01E
3-2 Appendix
2/4
QIS 12D06A03-01E
Table 1
Current Output (Indication)
Output Range
4.0 mA
4 ±0.02 mA DC
12.0 mA
12 ±0.02 mA DC
20.0 mA
20 ±0.02 mA DC
3.3 Temperature
Indication
Check
<Test>
x
From the last step in Item 3.2, Current Output Test, press the [ENT] key repeatedly until the
message display shows “PT1000.”
x
Change the resistance value of the resistance box 1 (for temperature) as shown in Table 2-1
and check the temperature readings. Each temperature reading must be within the range.
Table 2-1 (PT1000)
Resistance of
Resistance Box 1
Indication range
1097.3
:
25.0
±0.3°C
1385.0
:
100.0
±0.3°C
x
Press the [ENT] key. The message display will show “30k NTC.”
x
Change the resistance value of the resistance box 1 (for temperature) as shown in Table 2-2
and check the temperature readings. Each temperature reading must be within the range.
Table 2-2 (30k NTC)
Resistance of
Resistance Box 1
Indication range
30
k
:
25.0
±0.3°C
2.069
k
:
100.0
±0.3°C
x
From the above last step, proceed directly to the resistance (conductivity) indication check in
Item 3.4.
3.4 Resistance (Conductivity) Indication Check
<Test>
x
From the last step in Item 3.3, Temperature Indication Check, press the [ENT] key
repeatedly until the message display shows “SEL.10.”
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.1.”
x
Set resistance box 2 to the value shown in Table 3 and check the indication. The resistance
indication must be within the range.
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.2.” There is no need to check here, so just press the [ENT] key.
x
The message display shows “
WAIT
” momentarily, followed by “RES.3.”
x
Set resistance box 2 to the value shown in Table 3 and check the indication. The resistance
indication must be within the range.
x
Press the [ENT] key. The message display shows “
WAIT
” momentarily, followed by
“RES.4.” There is no need to check here, so just press the [ENT] key.
x
The message display shows “
WAIT
” momentarily, followed by “RES.5.”
Summary of Contents for vigilantplant EXA ISC202G
Page 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 12: ...IM 12D06A03 01E 1 4 Introduction...
Page 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D06A03 01E 2 20 Appendix...
Page 97: ...IM 12D06A03 01E Appendix 3 5...
Page 102: ...IM 12D06A03 01E 3 10 Appendix...
Page 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Page 106: ...IM 12D06A03 01E 3 14 Appendix...
Page 110: ...IM 12D06A03 01E 3 18 Appendix...