Sample 4
Concentration of hydrochloric acid
Meas. range
30-40%
Reference temp.
35°C
Temp. compensation
Enter temperature compensation coefficient,
Relate concentration to conductivity by table
START
Service code 20
Service code 31
Service code 35
Select 1
0% output --> 30 %
100% output --> 40%
Service code 55
TEMP
Calibration in Air
Calibration
35 °C (Default)
NaCl --> NO
TC --> YES ( here use [MODE] key to escape )
Calibrate conductivity value in actual operation
e.g. 34% HCl corresponds to conductivity 772 mS/cm
Select 1
Enter conductivity values corresponding to output current values.
Table (below) relates output current to conductivity at 35°C
Service code 21
TC=1.24 (data from customer)
Output
current (%)
0
20
40
60
80
100
HCl conc.(%)
30
32
34
36
38
40
Conductivity
(mS/cm)
845
840
772
731
688
644
sample4E.eps
Summary of Contents for vigilantplant EXA ISC202G
Page 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 12: ...IM 12D06A03 01E 1 4 Introduction...
Page 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D06A03 01E 2 20 Appendix...
Page 97: ...IM 12D06A03 01E Appendix 3 5...
Page 102: ...IM 12D06A03 01E 3 10 Appendix...
Page 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Page 106: ...IM 12D06A03 01E 3 14 Appendix...
Page 110: ...IM 12D06A03 01E 3 18 Appendix...