IM 12D06A03-01E
6-2 Calibration
ENT
NO
YES
MODE
MEASURE
CAL
DISPLAY
HOLD
MODE
6-2. Calibration procedure
Press the
MODE
key
The
legend
CALIB
appears,
and
the
YES/NO
key prompt flags flash
Put the sensor in standard solution
Press
YES
Set the value using the
key
Select the flashing digit with the
key
Increase it’s value by pressing the
key
When the correct value is displayed,
press
ENT
to enter the change
After briefly displaying
WAIT
,
the
CAL.END
message appears
The calibration is now complete
Put the sensor back in the process and press
YES
The cell constant is automatically updated after the calibration and the new value can be read on the dis-
lay as described in section 5-1-2. and 5-1-3.
If the calibrated cell constant is not within the range of 0.2 - 19.99 cm
-1
, error E3 is displayed.
Summary of Contents for vigilantplant EXA ISC202G
Page 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 12: ...IM 12D06A03 01E 1 4 Introduction...
Page 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D06A03 01E 2 20 Appendix...
Page 97: ...IM 12D06A03 01E Appendix 3 5...
Page 102: ...IM 12D06A03 01E 3 10 Appendix...
Page 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Page 106: ...IM 12D06A03 01E 3 14 Appendix...
Page 110: ...IM 12D06A03 01E 3 18 Appendix...