IM 12D06A03-01E
3-6 Installation and wiring
Extension cable may be purchased in bulk quantities, cut to length. Then it is necessary to terminate the
cable as shown below.
Termination procedure for WF10 cable.
1.
Slide 3 cm of heat shrink tube (9 x 1.5) over the cable end to be terminated.
2.
Strip 9 cm of the outer (black) insulating material, taking care not to cut or damage internal cores.
3 cm
heat shrink
9 cm
remove insulation
Fig. 3-10-1.
3.
Remove loose copper screening, and cut off the cotton packing threads as short as possible.
4.
Strip insulation from the last 3 cm of the brown, and the white coaxial cores.
Fig. 3-10-2.
5.
Extract the coaxial cores from the braid, and trim off the black (low-noise) screening material as short
as possible.
6.
Insulate the overall screen and drain wire (14) and the 2 coaxial screens with suitable plastic tubing.
7.
Strip and terminate all ends with suitable (crimp) terminals and identify with numbers as shown.
Fig. 3-10-3.
8.
Finally shrink the overall heat shrink tube into position.
Summary of Contents for vigilantplant EXA ISC202G
Page 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 12: ...IM 12D06A03 01E 1 4 Introduction...
Page 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D06A03 01E 2 20 Appendix...
Page 97: ...IM 12D06A03 01E Appendix 3 5...
Page 102: ...IM 12D06A03 01E 3 10 Appendix...
Page 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Page 106: ...IM 12D06A03 01E 3 14 Appendix...
Page 110: ...IM 12D06A03 01E 3 18 Appendix...