IM 12D06A03-01E
5-18 Parameter setting
5-3-5. User interface
Code 50
*RET.
When Auto return is enabled, the transmitter reverts to the measuring mode
from anywhere in the configuration menus, when no button is pressed during
the set time interval of 10 minutes.
Code 52
*PASS
Passcodes can be set on any or all of the access levels, to restrict access to
the instrument configuration.
Code 53
*Err.05 to Error message configuration. Two different types of failure mode can be set.
*Err.08
Hard fail gives a steady FAIL flag in the display. A fail signal is transmitted on
the mA output when enabled in code 32.
Soft fail gives a flashing FAIL flag in the display. A good example is the dry
sensor for a soft fail.
Code 54
*E5.LS
Limits can be set for shorted and open measurement. E5 (High) is default set
*E6.LIM
to 3 S and must be in the range of 0.10 S to 9.99 S. E6 (Low) is default set to
5
μ
S and must be in the range of 0.00
μ
S to 99.9
μ
S.
Code 55
*%
Linear weight percentage. For some applications the measured parameter
values may be (more or less) linear to concentration. For such applications it
is not needed to enter an output table, but 0 and 100% concentration values
directly can be set.
Code 56
*DISP.
The display resolution is default set to autoranging for conductivity reading. If a
fixed display reading is needed, a choice can be made out of 5 possibilities.
Code
Display
Function
Function detail
X
Y
Z
Default values
User interface
50
*RET.
Auto return
Auto return to measuring mode Off
Auto return to measuring mode On
0
1
1
On
51
Not used
52
*PASS
Passcode
Note # = 0 - 9, where
1=111, 2=333, 3=777
4=888, 5=123, 6=957
7=331, 8=546, 9=847
Maintenance passcode Off
Maintenance passcode On
Commissioning passcode Off
Commissioning passcode On
Service passcode Off
Service passcode On
0
#
0
#
0
#
0.0.0
Off
Off
Off
53
*Err.05
*Err.06
*Err.07
*Err.08
Error setting
Shorted measurement
Soft/Hard
Open measurement
Soft/Hard
Temperature sensor open Soft/Hard
Temp. sensor shorted
Soft/Hard
0/1
0/1
0/1
0/1
1
1
1
1
Hard
Hard
Hard
Hard
54
*E5.L S
*E6.LIM
E5 limit setting
E6 limit setting
Maximum conductivity value
Minimum conductivity value
3
5
S
μ
S
55
*%
*0%
*100%
Display mA in w/w%
mA-range displayed in w/w% off
mA-range displayed in w/w% on
Set 0% output value in w/w%
Set 100% output value in w/w%
0
1
0
Off
56
*DISP.
Display resolution
Auto ranging display
Display fixed to XXX.X
μ
S/cm
Display fixed to X.XXX mS/cm
Display fixed to XX.XX mS/cm
Display fixed to XXX.X mS/cm
Display fixed to XXXX mS/cm
0
3
4
5
6
7
0
Auto
57-59
Not used
Summary of Contents for vigilantplant EXA ISC202G
Page 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 12: ...IM 12D06A03 01E 1 4 Introduction...
Page 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D06A03 01E 2 20 Appendix...
Page 97: ...IM 12D06A03 01E Appendix 3 5...
Page 102: ...IM 12D06A03 01E 3 10 Appendix...
Page 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Page 106: ...IM 12D06A03 01E 3 14 Appendix...
Page 110: ...IM 12D06A03 01E 3 18 Appendix...