7. MAINTENANCE
7-1. Periodic maintenance for the EXA 202 transmitter
The EXA transmitter requires very little periodic maintenance. The housing is sealed to IP65 (NEMA 4X)
standards, and remains closed in normal operation. Users are required only to make sure the front win-
dow is kept clean in order to permit a clear view of the display and allow proper operation of the push-
buttons. If the window becomes soiled, clean it using a soft damp cloth or soft tissue. To deal with more
stubborn stains, a neutral detergent may be used.
NOTE:
Never used harsh chemicals or solvents. In the event that the window becomes heavily stained or
scratched, refer to the parts list (Section 10) for replacement part numbers.
When you must open the front cover and/or glands, make sure that the seals are clean and correctly fit-
ted when the unit is reassembled in order to maintain the housing’s weatherproof integrity against water
and water vapour. The measurement otherwise may be prone to problems caused by exposure of the
circuitry to condensation.
The EXA instrument contains a lithium cell to support the clock function when the power is switched
off. This cell needs to be replaced at 5 yearly intervals (or when discharged). Contact your nearest
Yokogawa service centre for spare parts and instructions.
7-2. Periodic maintenance of the sensor
Maintenance advice listed here is intentionally general in nature. Sensor maintenance is highly applica-
tion specific.
In general conductivity measurements do not need much periodic maintenance. If the EXA indicates an
error in the measurement or in the calibration, some action may be needed (ref. Section 8 troubleshoot-
ing).
Cleaning methods
1. For normal applications hot water with domestic washing-up liquid added will be effective.
2. For lime, hydroxides, etc., a 5 ...10% solution of hydrochloric acid is recommended.
3. Organic foulings (oils, fats, etc.) can be easily removed with acetone.
4. For algae bacteria or moulds, use a solution of domestic bleach (hypochlorite).
* Never use hydrochloric acid and bleaching liquid simultaneously. The very poisonous chlorine gas will
result.
Summary of Contents for vigilantplant EXA ISC202G
Page 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 12: ...IM 12D06A03 01E 1 4 Introduction...
Page 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D06A03 01E 2 20 Appendix...
Page 97: ...IM 12D06A03 01E Appendix 3 5...
Page 102: ...IM 12D06A03 01E 3 10 Appendix...
Page 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Page 106: ...IM 12D06A03 01E 3 14 Appendix...
Page 110: ...IM 12D06A03 01E 3 18 Appendix...