IM 12D06A03-01E
Parameter setting 5-15
Code Display
Function
Function
detail
X
Y
Z
Default
values
Temperature compensation functions
20
*T.R.°C
Set reference temp.
Use
keys to set value
25 °C
21
*T.C.1
Set temp. coef. 1
Adjust compensation factor
2.10 %
if set to *T.C. in section 5-2-5.
per °C
Set value with
keys
*T.C.2
Set temp. coef. 2
Adjust compensation factor
2.10 %
if set to *T.C. in section 5-2-5.
per °C
Set value with
keys
22
*MATRx
Select matrix
Choose matrix if set to *MATRX
in section 5-2-5, using
keys
H
2
SO
4
, 0 -100ºC, 0.5 - 5%
1
1
H
2
SO
4
H
2
SO
4
, 0 -100ºC, 2.5 - 25%
2
HCl, 0 - 60ºC, 0.5 - 5%
3
HCl, 0 - 60ºC, 1 - 20%
4
HNO
3
, 0 - 80ºC, 0.5 - 5%
5
HNO
3
, 0 -80ºC, 2.5 - 25%
6
NaOH, 0 -100ºC, 0.5 - 5%
7
NaOH, 0 -100ºC, 0.5 - 15%
8
User programmable matrix
9
23
*T1 °C (°F) Set temp. range
Enter 1st (lowest) matrix temp. value
*T2..
Enter 2nd matrix temp. value
*T3..
Enter 3rd matrix temp. value
*T4..
Enter 4th matrix temp. value
*T5..
Enter 5th (highest) matrix temp. value
24
*L1xT1
Enter conductivity
Value for T1
*L1xT2
values for lowest
Value for T2
....
concentration
*L1xT5
Value for T5
25
*L2xT1
Concentration 2
Similar to code 24
26
*L3xT1
Concentration 3
Similar to code 24
27
*L4xT1
Concentration 4
Similar to code 24
28
*L5xT1
Concentration 5
Similar to code 24
29
Not
used
Summary of Contents for vigilantplant EXA ISC202G
Page 8: ...IM 12D06A03 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 12: ...IM 12D06A03 01E 1 4 Introduction...
Page 68: ...IM 12D06A03 01E Spare parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D06A03 01E 2 20 Appendix...
Page 97: ...IM 12D06A03 01E Appendix 3 5...
Page 102: ...IM 12D06A03 01E 3 10 Appendix...
Page 105: ...IM 12D06A03 01E Appendix 3 13 3 3 QIS 12D06A03 61E Figure 1 Testing Circuit and Test Equipment...
Page 106: ...IM 12D06A03 01E 3 14 Appendix...
Page 110: ...IM 12D06A03 01E 3 18 Appendix...