52
IM MY600-01EN
4.
With the measurement switch locked or the remote
switch held down, press 0Ω ADJ.
lights, and 0.00 Ω is displayed. The value will be
stored in memory and will not be cleared even when
the power is turned off.
5.
To clear the value, press 0Ω ADJ with the probe
opened. When the value is cleared,
disappears.
When the resistance cannot be cancelled, “no” will be
displayed, and 0Ω adjustment will not be executed even
if you press 0Ω ADJ. If you want to cancel the resistance,
contact your nearest YOKOGAWA dealer to have the
instrument adjusted.
Measurement Procedure
1.
Set the range switch to the V/Ω position.
2.
If the instrument is set to voltage measurement
mode, press SELECT to change to low-resistance
measurement mode.
3.
Connect the probe to the DUT, and press the
measurement switch or remote switch.
Measurement principle:
Resistance = Voltage/Current
RX = V/ I
Voltage: V
LINE(+)
EARTH(-)
Current: I
Resistance: RX
• If a current of 200 mA or more runs during the
measurement, a buzzer sounds to indicate that power
is running. You can turn off this buzzer (see chapter 8,
“Backlight, LED Light, and Buzzer”).
• The measurement results of low-resistance
measurement may be affected by the impedance of the
operating circuit connected in parallel.