• To prevent mistakes in operation, an empty range is
provided between the 500 V and 1000 V ranges of the
range switch.
• Intermittent warning buzzer sounds when the range
switch is set to 1000 V range.
• Hold function
A measured insulation or low-resistance value is held
and displayed until the range is changed or another
measurement is started.
• 0Ω adjustment function
In low-resistance measurements, the resistance of the
cord and fuse can be cancelled up to about 3 Ω.
• Elapsed time display
The measurement time is displayed upon starting an
insulation resistance measurement.
• One-minute value display
The measurement 1 minute after starting an insulation
resistance measurement can be displayed.
• Dielectric Absorption Ratio (DAR)/Polarization Index (PI)
measurement
DAR and PI values are calculated automatically during
an insulation resistance measurement.
• Memory function
Measured data can be saved, loaded, and deleted from
the internal memory.
• Communication function
Data saved in the internal memory can be transferred to
a PC using communication.
• Clock
Measured data is saved with the measurement time
information.