Continuous Measurement
To perform insulation resistance measurement
continuously, hold down the measurement switch and turn
it clockwise. The measurement switch will be locked. Then,
press the probe’s remote switch to measure continuously.
If you change the range, save data to memory, or perform
any non-measurement operation during a continuous
measurement, you will need to turn off the measurement
switch and the probe’s remote switch once in order to
make another measurement.
When the measurement is complete, turn the measurement
switch counterclockwise to return it to its original position.
WARNING
High voltage is generated continuously at the tip of the
probe. Be careful of electric shock.
French
AVERTISSEMENT
Ne pas toucher les pointes des fils d’essais pour éviter les
chocs électriques, car une haute tension est présente en
permanence.
Voltage Characteristics of the Signal
Terminals
The instrument complies with JISC1302. This standard
defines that the rated measurement current must be at
least 1 mA. It also specifies the lower limit of the insulation
resistance for maintaining the rated measurement voltage
at measurement terminals. (See the table below.)