23
IM MY600-01EN
2. Specifications
Measuring range and accuracy
23°C±5°C, RH 80% or less
Accuracy: Within 1 year of shipment
Voltage Measurement
Range
300.0/600 V (auto range)
Display range
300.0 V: 0.0 to 314.9 V
600 V: 270 to 629 V
Measuring range
(accuracy guaranteed)
AC: 2.0 to 600 V rms (45 to 65 Hz)
DC: ±2.0 to ±600 V
Over-range display
AC: >629 V
DC positive: >629 V
DC negative: < -629 V
Accuracy
±1%rdg±4dgt
* RMS detection in AC mode. For sine waves other than with
CF<2.5, add ±1%rdg for the above listed accuracy.
AC/ DC auto-detection (2 V or higher)
Low-resistance measurement (Continuity check)
Resistance range
40.00/400.0/4000 Ω (auto range)
Open-circuit voltage (DC) 5 V (4 to 6.9 V)
Measurement current
200 mA or more (2 Ω or less)
Display range
40.00 Ω: 0.00 to 41.99 Ω
400.0 Ω: 36.0 to 419.9 Ω
4000 Ω: 360 to 4199 Ω
Over-range display
>4199 Ω
Measuring range and
accuracy
0.20 to 4000 Ω (to
keep operating
uncertainty within
limits)
0 to 0.19 Ω
±2.5%rdg ±8dgt
±8dgt