A2-2
IM 34M06H63-01E
2nd Edition:Jul.16,2015-00
A2.3 General Specifications
Table A2.2 lists the general specifications of the F3CX04-0N temperature monitoring
module.
Table A2.2 General Specifications
Item Specification
Number of channels
4
Isolation
Between input terminals and internal circuit
Isolated by photocouplers and transformers
(tested for 1500 V AC voltage withstanding)
Between input terminals
Alarm types
4 types of alarm:
input upper limit alarm, input lower limit alarm
(with or without waiting for each of the above 2 alarms)
Number of alarm outputs (input relays)
4 points per channel (Only alarms 1 and 2 have input relays)
Alarm delay timer
Yes
Warm-up time
30 minutes min.
Max. allowable ambient temperature change rate
*1
10
C/h max.
Mounting position
Horizontal or inverted orientation not allowed
External connection
18-point terminal block with M3.5 screws
External dimensions
* 2
28.9 (W) x 100 (H) x 106.1 (D) mm
Current consumption
440 mA at 5 V DC
Weight 200
g
*1: The stated accuracy for the reference junction for thermocouple input deteriorates if the ambient temperature change exceeds this rate.
*2: External dimensions excluding protrusions (for details, see the External Dimensions drawing).
A2.4 Input Specifications
Table A2.3 lists the input specifications of the F3CX04-0N temperature monitoring
module.
Table A2.3 Input Specifications
Item
Specification
Input sampling period
*1
200 ms for 4 channels, or 100 ms for 2 channels
Input types and ranges
See Table A2.4, “Instrument Range and Accuracy”.
Individual inputs separately configurable by software or collectively by
hardware
Thermocouple input : 15 ranges
RTD input
: 9 ranges
DC voltage input
: 6 ranges
Burnout detection
hermocouples or RTDs are checked for burnout.
Up-scale, down-scale, or none may be selected.
Detection
current
Thermocouple
100 nA max.
RTD
100 nA max.
Input insulation resistance
1 M
min.
Allowable signal
source resistance
Thermocouple or DC mV input
250
max.
DC voltage input
2 k
max.
Allowable wiring
resistance
RTD 10
max. per wire (three wires must have the same resistance)
Measuring current
RTD
Approx. 270
μ
A
Reference junction
compensation
Thermocouple
*2
2.0
C (0 to 55
C)
Allowable input voltage range
-20 to 20 V DC
Noise reduction
*3*4
Common mode
120 dB (50/60 Hz) min.
Normal mode
40 dB (50/60 Hz) min.
Effect of ambient temperature
0.01%/
C or
1
μ
V/
C, whichever is greater
*1: If input sampling period is set to 100 ms for 2 channels, only CH1 and CH2 are available.
*2: This value assumes that all input terminals are correctly wired (that is, solderless termination, wire diameters and connections are correct).
*3: This value assumes that the power supply frequency is correctly selected.
*4: This module continues to operate at a input accuracy of ±0.5% max. of F.S. during the radiated electromagnetic field test.
Summary of Contents for FA-M3 VITESSE F3CX04-0N
Page 2: ...Blank Page...
Page 16: ...Blank Page...
Page 18: ...Blank Page...
Page 32: ...Blank Page...
Page 44: ...Blank Page...
Page 68: ...Blank Page...
Page 80: ...Blank Page...
Page 82: ...Blank Page...
Page 90: ...Blank Page...
Page 110: ...Blank Page...
Page 118: ...Blank Page...
Page 126: ...Blank Page...
Page 128: ...Blank Page...
Page 132: ...Blank Page...
Page 142: ...Blank Page...
Page 144: ...Blank Page...