![YOKOGAWA DpharpEJX vigilantplant EJX910A User Manual Download Page 62](http://html1.mh-extra.com/html/yokogawa/dpharpejx-vigilantplant-ejx910a/dpharpejx-vigilantplant-ejx910a_user-manual_3750163062.webp)
IM 01C25R01-01E
10-7
10. GENERAL SPECIFICATIONS
10.3 OPTIONAL SPECIFICATIONS
T1012E.eps
Code
FF1
FM Explosionproof Approval
Explosionproof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
in Hazardous locations, indoors and outdoors (NEMA 4X)
Temperature class: T6, Amb. Temp.: –40 to 60
#
C (–40 to 140
#
F)
Description
Item
Factory Mutual (FM)
CENELEC ATEX (KEMA) Intrinsically safe Approval
*2
Applicable Standard: EN 50014, EN 50020, EN 50284, EN 50281-1-1
Certificate: KEMA 06ATEX0037X II 1G, 1D EEx ia IIC T4 Type of protection : IP67
Amb. Temp. (Tamb) for gas-proof: –50 to 60
°
C (–58 to 140
°
F)
Maximum Process Temp.(Tp) for gas-proof: 120
°
C
Electrical data:
[Supply/Output circuit (ter and –)] Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH
[Pulse Output circuit (terminals – and pulse)] Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH
[External Temperature Input circuit (connector)] Uo=30 V, Io=95.4 mA, Po=468 mW, Co=11 nF,
Lo=3.9 mH
Max. surface Temp. for dust-proof: T85
°
C (Tamb: –40 to 60
°
C, Tp: 80
°
C),
T100
°
C (Tamb: –40 to 60
°
C, Tp: 100
°
C), T120
°
C (Tamb: –40 to 60
°
C, Tp: 120
°
C)
Combined FF1 and FS1
CSA Explosionproof Approval
*1
[For CSA C22.2]
Explosion-proof for Class I, Groups B, C and D.
Dustignition-proof for Class II/III, Groups E, F and G.
When installed in Division 2, “SEAL NOT REQUIRED” Enclosure: TYPE 4X, Temp. Code: T6...T4
[For CSA E60079]
Flameproof for Zone 1, Ex d IIC T6...T4 Enclosure: IP66 and IP67
Max.Process Temp.: T4; 120
#
C (248
#
F), T5; 100
#
C (212
#
F), T6; 85
#
C (185
#
F)
Amb.Temp.: –50 to 75
#
C (–58 to 167
#
F) for T4, –50 to 80
#
C (–58 to 176
#
F) for T5,
–50 to 70
#
C (–58 to 158
#
F) for T6
CSA Intrinsically safe Approval
*1
Combined KF2, KS2 and Type n
*1
CENELEC ATEX (KEMA) Flameproof Approval
II 2G,1D EExd IIC T4, T5, T6
Amb. Temp. (Tamb) for gas-proof: T4; –50 to 75
°
C (–58 to 167
°
F),
T5; –50 to 80
#
C (–58 to 176
#
F), T6; –50 to 70
°
C (–58 to 158
°
F)
Max. process Temp.(Tp): T4; 120
#
C (248
#
F), T5; 100
#
C (212
#
F), T6; 85
#
C (185
#
F)
Max. surface Temp. for dust-proof: T80
°
C (Tamb: –40 to 40
#
C, Tp: 80
#
C),
T100
°
C (Tamb: –40 to 60
#
C, Tp: 100
#
C), T120
°
C (Tamb: –40 to 80
#
C, Tp: 120
#
C)
Type of protection: IP66 and IP67
FM Intrinsically safe Approval
*1
—
—
CF1
KS2
—
KF2
—
CENELEC ATEX
—
Canadian Standards
Association (CSA)
Combined CF1 and CS1
*1
Contact Yokogawa representative for the codes indicated as ‘—’
*1:
Applicable for electrical connection code 2 and 7.