![YOKOGAWA DpharpEJX vigilantplant EJX910A User Manual Download Page 24](http://html1.mh-extra.com/html/yokogawa/dpharpejx-vigilantplant-ejx910a/dpharpejx-vigilantplant-ejx910a_user-manual_3750163024.webp)
IM 01C25R01-01E
3-8
3. HANDLING CAUTIONS
3.12Low Voltage Directive
Applicable standard : EN61010-1
(1) Pollution Degree 2
“Pollution degree” describes the degree to which a
soild, liquid, or gas which deteriorates dielectric
strength or surface resistivity is adhering. “2”
applies to normal indoor atmosphere. Normally,
only non-conductive pollution occurs. Occasionally,
however, temporary conductivity caused by
condenstaion must be expected.
(2) Installation Category I
“Overvoltage category(Installation category)”
describes a number which defines a transient
overvoltage condition. It implies the regulattion for
impulse withstand voltage. ” I ” applies to electrical
equipment which is supplied from the circuit when
appropriate transient overvoltage control means
(interfaces) are provided.