IM 01C25R03-01E
9-3
9. PARAMETER LISTS
Write
Mode
Explanation
Factory
Default
Relative
Index
Parameter Name
Index
DEVICE_STATUS_2
1046
1047
1048
1049
1050
1051
1052
–
–
–
–
–
–
–
AUTO
AUTO
0
Device status For details, refer to Table 8.2
46
DEVICE_STATUS_3
0
Device status For details, refer to Table 8.3
47
DEVICE_STATUS_4
0
Device status For details, refer to Table 8.4
48
DEVICE_STATUS_5
0
Device status For details, refer to Table 8.5
49
DEVICE_STATUS_6
0
Device status For details, refer to Table 8.6
50
DEVICE_STATUS_7
0
Device status For details, refer to Table 8.7
51
DEVICE_STATUS_8
0
Device status For details, refer to Table 8.8
52
1053
1054
1055
1056
1057
1058
–
–
–
–
SOFTDWN_PROTECT
0x01
Defines whether to accept software downloads.
0x01: Unprotected
0x02: Protected
53
SOFTDWN_FORMAT
0x01
Selects the software download method.
0x01: Standard
54
SOFTDWN_COUNT
0
Indicates the number of times the internal FlashROM was erased.
55
SOFTDWN_ACT_
AREA
0
Indicates the ROM number of the currently working FlashROM.
0: FlashROM #0 working
1: FlashROM #1 working
56
SOFTDWN_MOD_REV
1, 0, 0, 0, 0, 0, 0, 0, 0
Indicates the software module revision.
57
SOFTDWN_ERROR
0
Indicates the error during a software download.
Refer Table A7.4.
58
T0901-3.EPS
9.2
SENSOR Transducer Block
T0902-1.EPS
Write
Mode
Explanation
Factory
Default
Relative
Index
Parameter Name
Index
2000
Block Header
Block Tag
= O/S
TAG: “STB”
0
Information on this block such as Block Tag, DD Revision,
Execution Time etc.
2001
ST_REV
–
–
1
The revision level of the static data associated with the
function block. The revision value will be incremented each
time a static parameter value in the block is changed.
2002
TAG_DESC
AUTO
Null
2
The user description of the intended application of the block
2003
STRATEGY
AUTO
1
3
The strategy field can be used to identify grouping of blocks.
This data is not checked or processed by the block.
2004
ALERT_KEY
AUTO
1
4
The identification number of the plant unit. This information
may be used in the host for sorting alarms, etc.
2005
MODE_BLK
AUTO
AUTO
5
The actual, target, permitted, and normal modes of the block.
2006
BLOCK_ERR
–
–
6
This parameter reflects the error status associated with
hardware or software components associated with a block. It
is a bit string, so that multiple errors may be shown.
2007
UPDATE_EVT
–
–
7
This alert is generated by any change to the static data.
2008
BLOCK_ALM
–
–
8
The block alarm is used for all configuration, hardware,
connection failure or system problems in the block. The cause
of the alert is entered in the subcode field. The first alert to
become active will set the Active status in the Status attribute.
2009
TRANSDUCER_
DIRECTORY
–
–
9
A directory that specifies the number and starting indices of
the transducers.
2010
TRANSDUCER_
TYPE
–
100 (Standard Pre-
ssure with Calibration)
10
Identifies transducer.
2011
XD_ERROR
–
–
11
The error code in transducer.
0=No failure
18=Calibration error
20=Electronics failure
22=I/O failure
2012
COLLECTION_
DIRECTORY
–
–
12
A directory that specifies the number, starting indices, and
DD Item Ids of the data collections in each transducer within
a transducer block.